111
HistoCore SPECTRA CV
Cleaning and Maintenance
7
7.2.9 Removing the complete unit of the needle cleaning container
The needle cleaning container is inserted into a red complete unit
, which can be removed
by the user through the opening in the unload drawer. The complete unit is hooked into the coverslip
line by two lateral pins
. The removal may be necessary if the needle cleaning container
is stuck due to dried mounting medium and cannot be removed as described
6.3.2 Monitoring and refilling of the needle cleaning container)
1
2
3
Fig. 88
Warning
•
There is a danger of cutting injuries when removing the complete unit. Therefore, proceed with
the necessary level of caution and wear cut-resistant gloves
P. 18 – 3.1 Standard delivery –
.
•
P. 32 – 4.6 Refilling consumables)
must be observed when
handling solvents!
1. Open the hood.
2. Call up the
Module Status
and press the
Prime/Clean
button of the respective
L1
or
L2
coverslip
line.
3. The needle moves upwards out of the needle cleaning container.
4. Remove the needle from the holder and insert it into the prime bottle.
5. Open the load and unload drawers completely and remove all racks.