Daily Instrument Setup
6
1
2
3
4
5
Fig. 59
6.3.3 Checking and replacing the coverglass cartridge
Warning
While changing the coverglass cartridge, always wear personal protective clothing (cut-resistant
gloves, safety goggles)!
Note
•
For the HistoCore SPECTRA CV Leica-validated coverglass
are
available. The coverglasses are only available in cartridges which are equipped with a RFID chip.
Information on the cartridge used (e.g. quantity and size) is automatically read when it is inserted
and the hood is closed.
•
Changing the coverglass cartridge in the
L2
coverslip line is described in the following. The same
procedures also apply for the coverslip line
L1
.
•
For each newly inserted rack, the consumable management system (CMS) calculates whether the
available quantity of coverglass in the magazine is sufficient.
•
If the CMS determines that a newly inserted rack can no longer be completely coverslipped, an
appropriate warning message
will be displayed to the user.
•
If there is a rack located in the load drawer, or if one is to be placed there, the CMS indicates to the
user that a new coverglass cartridge must be inserted