127
HistoCore SPECTRA CV
Cleaning and Maintenance
7
9.
The instrument software identifies that no mounting medium bottles, coverglass cartridges and
needle cleaning containers are inserted, and indicates this to the user through error messages
for each coverslip line. Confirm these messages with
OK
.
10. Finally, shut down the instrument, switch it off at the
power switch
and disconnect it from the
power grid.
The instrument is now ready for transport or for storage.
7.3.4 Recommissioning after transport or storage
Note
•
For recommissioning after a transport or after long-term storage, proceed as described in
P. 24 – 4. Installation and Instrument Setup)
and
P. 75 – 6. Daily Instrument Setup)
•
The internal battery
P. 26 – 4.2.1 Internal battery)
has to be charged if the instrument has
been disconnected from a power supply for 4 weeks. For this purpose, connect the instrument to
the socket and switch on the
Main switch
. The charging time is approx. 2 hours.
7.4 Recommended cleaning and maintenance intervals
Warning
•
It is mandatory to observe and carry out the following cleaning and maintenance intervals and
cleaning tasks.
•
In case of non-compliance with the maintenance intervals, the proper function of the
HistoCore SPECTRA CV by Leica Biosystems GmbH cannot be guaranteed.
• The instrument should be inspected once per year by a qualified service technician authorized by
Leica.
•
Always wear suitable protective clothing (lab coat, cut-resistant gloves, safety goggles).
To ensure trouble-free operation of the instrument over a long period of time, it is strongly recommended
that you conclude a maintenance contract after the warranty period has elapsed. For more information,
contact the relevant customer service organization.