HistoCore SPECTRA CV
Installation and Instrument Setup
4
4.6.3
Prepare the reagent vessel, fill it and insert it into the load drawer
To ensure that everything runs smoothly in the instrument, observe the following notes and proceed as
follows.
Attaching the reagent vessel handle:
» Check whether the reagent vessel handle is correctly attached to the reagent vessel. If this is not
the case, fasten the handle according to
.
Fig. 17
1
2
Correct filling of reagent vessels:
Note
• The reagent vessel must be filled with xylene
P. 32 – 4.6 Refilling consumables)
• Reagent vessels must always be filled outside the instrument.
•
Make sure that the reagent vessel handle is not bent. During the removal of the reagent vessel, if
the handle is bent out of shape, it may detach from the reagent vessel and the reagent could be
spilled.
Warning
The safety notes for reagent handling must be observed!