80
Version 2.1, Revision K
Daily Instrument Setup
6
Fig. 57
1
2
6. Insert the new mounting medium bottle
into the
L2
bottle sledge and make sure that the RFID chip attached to the bottle is seated in the notch of
7. Take the cannula with the label
2
and
carefully insert into the opening of the mounting medium bottle
until it noticeably
clicks into place
. In the course of this, the white protective membrane of the
bottle is punctured.
Note
Before inserting the drain tube into the new mounting medium bottle, check the cannula for any dried
mounting medium residue and, if necessary, moisten with xylene and clean with a lint-free cloth.
Fig. 58
1
6
2
7
3
4
5
Warning
Puncturing the white protective membrane briefly requires an increase in force. Therefore, proceed
carefully to avoid letting it slip from your hand and to avoid bending the cannula.
back into its original position once again until it clicks into
place.