34
Version 2.1, Revision K
Installation and Instrument Setup
4
1
2
Fig. 14
4.6.2 Inserting the mounting medium bottle and prime bottle
1. Open the hood.
2. Grasp the bottle sledge on the handle
and move it all the way forward.
Inserting the prime bottle
» Insert the prime bottle
into the position provided for it
sledge.
Inserting the mounting medium bottle
Note
•
Before the insertion of a mounting medium bottle in the instrument, the expiration date
(imprint on the packaging and on the bottle label
) is to be checked. If the expiration
date has been reached or exceeded, the mounting medium may no longer be used. Expired
mounting medium is to be disposed of in accordance with the applicable on-site lab guidelines.
•
Opened mounting medium has a storage life of 14 days, within which it must be used up.
We recommend making a note of the use-date on the bottle label.
•
Only original mounting medium that has been validated by Leica may be used
in order to prevent instrument faults.
Warning
Ensure the correct assignment of the cannulas
to the respective mounting medium
bottles to prevent instrument faults. The cannula with the mark 1 is to be inserted into the mounting
medium bottle in position 1 and the cannula with the mark 2 into the mounting medium bottle in
position 2 of the bottle sledge
.