131
HistoCore SPECTRA CV
Malfunctions and Troubleshooting
8
8. Malfunctions and Troubleshooting
8.1 Troubleshooting
Problem/malfunction
Possible cause
Correction
Black screen
•
Software crash
•
Remove all racks from the
instrument and restart the
instrument.
•
Power failure
•
scenario and instrument failure)
Needle cleaning container
cannot be removed using
the knurled screw
•
Needle cleaning container is
stuck due to mounting medium
residue in the holder.
•
Remove the needle cleaning unit
and immerse it in xylene until the
needle cleaning container has
loosened and can be taken out
the complete unit of the needle
Consumables cannot be
scanned
•
RFID chip is dirty
•
Carefully remove any
contamination from the RFID chip
and re-scan the consumables by
closing the hood.
•
RFID chip is defective
•
Try again with new consumables.
•
There may be an instrument defect
present if the problem persists.
Please contact Leica Service.
Load/unload drawer is
sticking
•
Possible contamination in or on
the respective drawer
•
Carefully remove any
contamination (mounting
medium residue, glass splinters)
, remove all racks from
the instrument and restart the
instrument.
Transport arm reports
an error
•
Malfunction in the instrument
•
Follow the instructions in
removal of a rack in the event of
and pull
the transport arm to the unload
drawer. Remove the rack from the
transport arm.