141
HistoCore SPECTRA CV
Malfunctions and Troubleshooting
8
1
2
3
Fig. 111
5. Check the coverslip line for broken glass and clean if necessary.
6. If the
Pick&Place module
was unable to remove a coverglass from the coverglass cartridge,
check the suction cups on the
Pick&Place module
for contamination or damage and, if
necessary, clean or replace the suction cups with two new ones.
7.
Then check the fill level of the coverglass cartridge as well. If there is no longer enough
coverglass present, insert a new coverglass cartridge.
8. If the steps listed above have been carried out, press the
OK
button and close the hood.
9.
The instrument software performs a fill level scan. The user receives an information message
.
10. If the processing was interrupted for too long, it can be canceled by pressing
Yes
. The rack in question is transported to the unload drawer and can be removed by
the user there.
11. To continue processing, press the
No
button. Processing is continued as normal.