Cleaning and Maintenance
7
7.4.3 Quarterly cleaning and maintenance
A
Changing the active carbon filter:
• Replace the active carbon filter
P. 117 – 7.2.15 Changing the active carbon filter)
.
7.4.4 Cleaning and maintenance as necessary
Warning
•
Attention: Wear cut-resistant gloves
P. 18 – 3.1 Standard delivery – packing list)
•
Clean the interior of the instrument in the event of broken glass.
•
A lint-free cloth moistened with xylene can be used to clean off any spilled mounting media.
You can use the brush included in the delivery package to brush glass debris without residue
of mounting media or loose glass debris toward the drawer outlet and then use a commercially
available vacuum cleaner to carefully clean them up.
A
•
Make sure the instrument is idle
P. 76 – 6.2 Switching on and shutting down the
:
•
Cover reagent vessels
•
Ensure there is enough xylene in the needle cleaning container to prevent the needle from
drying out
P. 81 – 6.3.2 Monitoring and refilling of the needle cleaning container)
B
•
Prime the hose system with xylene
P. 123 – 7.3.3 Cleaning the hose system)
.
C
Pick&Place module:
•
Replace the suction cups if they have severe deformation or are very dirty
D
•
Cleaning the interior of the instrument
P. 107 – 7.2.4 Interior cleaning)
E
To ensure a smooth operation of the instrument software, the instrument must be restarted
at least every 3 days