Cleaning and Maintenance
7
Cleaning process finished and preparation for transport or storage
1. After the cleaning process has been completed, remove the prime bottle from the instrument
and dispose of the contents according to laboratory specifications.
2. Insert the empty prime bottle back into the instrument again.
Note
After the cleaning process has been completed, the user is asked if a new bottle of mounting medium
is to be used to continue with the coverslipping
.
3. If the instrument is to be prepared for transport or storage, press the
No
button.
4. Take note of the next information message
, which contains instructions, and
acknowledge with the
Ok
button.
Fig. 105
5. After pressing the
OK
button, the instrument immediately begins pumping air through the hose
system.
6. On the screen, an information message
appears informing the user about the
duration. This message disappears as soon as air pumping has been completed.
Fig. 106
7. Repeat the procedure with the
L2
coverslip line.
8. After the air pumping has been completed for both coverslip lines, remove the remaining
coverglass cartridges and needle cleaning containers and close the instrument hood.