137
HistoCore SPECTRA CV
Malfunctions and Troubleshooting
8
4. The instrument is then shut down by the software in a controlled manner
P. 30 – 4.5 Switching on and shutting down the instrument)
L
After power supply has been restored, the instrument restarts automatically. An audible warning
signal sounds and a warning message is displayed on the display
Fig. 108
5.
Confirm this message with
OK
, whereupon the instrument shuts down in a controlled manner.
Then follow the instructions in the information message
and remove the slides from the
instrument.
Note
•
After the restart of the instrument, the oven is shut down, and the user must reactivate it in the
Oven Settings menu
. The user is notified by a information
message upon the setting of a rack.
•
For a safe and easy removal of the racks from the instrument, the instrument must be shut down.
Following shutting down, the user can, for example, easily move the transport arm in order to carry
out the necessary steps for removing the racks.
•
Check the removed racks for un-coverslipped slides and coverslip those slides manually. In doing
so, observe the sequence in which the individual slides are processed in the coverslipping process
P. 94 – 6.6 Starting the coverslipping operation)
and
.
•
The following chapters shall be observed for removal of the rack remaining in the instrument.
Position
See
Load drawer
P. 138 – 8.3 Manual removal of a rack in the event of instrument
Rotator
P. 150 – 8.3.5 Removal of the rack from the rotator)
Elevator
P. 142 – 8.3.2 Removing a rack from the elevator of the coverslip line)
Transport arm
P. 138 – 8.3 Manual removal of a rack in the event of instrument
Oven
8.3.4 Removing the rack from the oven or from behind the oven)
Unload drawer
P. 138 – 8.3 Manual removal of a rack in the event of instrument