19
HistoCore SPECTRA CV
Instrument Components and Specifications
3
Note
The delivered components must be carefully compared against the packing list, delivery note, and your
order. Should you find any discrepancies, please contact your Leica sales office without delay.
3.2 Technical data
Nominal supply voltages:
100–240 V AC
+
10 %
Nominal frequency:
50/60 Hz
Power draw:
1100 VA
Fuses:
2 x T16 A H 250 V AC
IEC 1010 classification:
Protection class 1
Pollution degree in accordance with IEC61010-1:
2
Overvoltage category in accordance with IEC61010-1:
II
Exhaust air:
Hose length:
2000 mm
Inner diameter:
50 mm
Outer diameter:
60 mm
Exhaust performance:
30 m
3
/h
Exhaust extraction:
Active carbon filter and exhaust hose for connecting with an external
exhaust system.
Heat emission:
1100 J/s
A-weighted noise level, measured at 1 m distance:
< 70 dB (A)
Connections:
1 x RJ45 Ethernet (rear):
RJ45 - LAN (external data management)
1 x RJ45 Ethernet (front):
Only for service purposes
2 x USB 2.0 (front):
5 V/500 mA (service & data storage)
International protection class:
1st parameter = Protected against fixed foreign bodies with a
diameter ≥ 12.5 mm
2nd parameter = No protection against water
IP20
Ambient conditions:
Operation:
Temperature:
+18 °C to +30 °C
Relative humidity:
20 % to 80 %, non-condensing
Operating elevation:
Up to a max. of 2000 m above sea
level
Storage:
Temperature:
+5 °C to +50 °C
Relative humidity:
10 % to 85 %, non-condensing
Transport:
Temperature:
-29 °C to +50 °C
Relative humidity:
10 % to 85 %, non-condensing