96
Version 2.1, Revision K
Daily Instrument Setup
6
Note
•
The slides are not processed sequentially, i.e. they are not coverslipped in sequence.
•
The schematic display
of a rack
shows the sequence during processing.
• The first five slides are coverslipped at the beginning
. Then the rack moves upwards
and continues the coverslipping with the lowest slide
. The sixth slide (counted from
above) is coverslipped last. This procedure prevents the lowest slide from drying out.
•
Finished coverslipped slides are moved back to their original position within the rack.
•
The instrument detects and indicates when a rack has been inserted in the wrong orientation and
has to be corrected by the user.
•
The system detects if a rack has been inserted into the load drawer with a rack handle color for
which no parameter set is active, and the user is informed by a corresponding message. The rack
must be removed from the instrument. Corresponding to the startable parameter sets
, attach the correct colored handle to the rack and re-insert it into the
load drawer.
•
Racks that are inserted into the left-side reagent vessel in the load drawer are coverslipped in the
L1
coverslip line. Racks from the right-side vessel are coverslipped in the
L2
coverslip line.
1
2
3
4
Fig. 74