55
HistoCore SPECTRA CV
Operation
5
•
The volume setting for error sounds is adjusted using the
Volume
roller. Six different sounds are
available for the alarm types. The
Test
button can be used to listen to the respective
sound.
•
The volume can be adjusted incrementally by turning the roller.
Note
The acoustic error sounds cannot be disabled. The minimum configurable value for the volume is 2.
The maximum value is 9.
•
Pushing the
Save
button saves the settings. The
Cancel
button is used to
close the selection window without applying the settings.
•
Press the
Back
button to return to the
Settings
menu.
5.8.5 Oven settings
Required access level: Standard user, supervisor
The operating mode of the oven can be configured in the menu for oven settings
. The current oven settings are displayed after calling up the menu.
Note
•
When starting the instrument, the oven is normally enabled and the
Exclude Oven Step
• The oven heats up to approx. 35 °C. As soon as a rack has been coverslipped, the oven continues
heating up to approx. 40 °C. The symbol in the status bar changes from the status "heating up"
to the status "ready to operate"
.
•
The oven dries the cover slipped slides after completed processing for approx. 5 minutes. After the
oven step has ended, the mounting medium is not completely dry. Treat the slides carefully during
the removal from the rack to avoid a sliding of the coverglass.
• The temperature setting of the oven (maximum 40 °C) and the dwell time of the finished
coverslipped slides are both configured at the factory and cannot be changed by the user.
• It can take up to 4 minutes for the oven to heat up!