Cleaning and Maintenance
7
4. Open the instrument hood and remove the needle
from the holder
).
5. Stick the needle unit
information message
with
OK
.
Warning
Do not close the hood while the needle is inserted in the prime bottle in order to avoid kinking the hose
or bending the needle.
Fig. 97
1
2
1
3
In the following menu
, three different priming/cleaning programs are available.