HistoCore SPECTRA CV
Daily Instrument Setup
6
Warning
•
The front side of the rack is marked with the
Leica
logo
. With the handle attached,
Front
can be read when looking at the handle
.
•
When inserting the slides, be absolutely certain that the slide label is pointing upwards and toward
the user
. The side of the slide with the specimen must point toward the front side of
the rack.
•
If the slides are not inserted correctly, the samples can become damaged in the subsequent
processing.
•
The stickers and/or labels used on the slides must be resistant to the solvents (xylene) used in the
instrument.
•
Labels that are used must not reach into the coverglass support area, and should not overhang
over the edge of the slide, in order to avoid specimen impairment or instrument fault during
coverslipping.
Fig. 70
1
1
3
2
Note
•
When using printable labels or manual labels on the rack handles and the slides, they must be
checked for solvent (xylene) resistance prior to use.
•
Make sure that only one slide each is put into the slot and that the slides are not jammed. Slots
must not be empty between the two slides.