N9030B PXA Signal Analyzer Service Guide 127
Instrument Messages
Condition Messages
Condition Errors 67 to 95, Calibration Failure (Extended)
This series of errors corresponds to the bits in the
STATus:QUEStionable:CALibration:EXTended:FAILure sub-register. The second
column in the table shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 71 indicates that the Characterize Preselector Failure has
been detected, error 1071 indicates that failure has been cleared.
This register is summarized as bit 9 of the STATus:QUEStionable:CALibration
register. See section
“Condition Errors 36 to 64, Calibration Needed or Failed”
Err#
Bit in
status
register
Message
Error or
Warning
More Information
67
0
Align 9kHz to 30MHz
failed
W
On PSA, this was error 13749
69
1
Align 30MHz to 1GHz
failed
W
On PSA, this was error 13751
71
2
Characterize Preselector
failure
W
The preselector characterization routine
failed.
73
3
unused
75
4
unused
77
5
unused
79
6
unused
81
7
unused
83
8
unused
85
9
unused
87
10
unused
89
11
unused
91
12
unused
93
13
unused
95
14
unused
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...