N9030B PXA Signal Analyzer Service Guide 261
LO Synthesizer/Reference Troubleshooting
A14 LO Synthesizer and A20 YTO Troubleshooting
7.
Verify the 100 MHz signal from the A16 Reference Assembly is stable by
pressing
SPAN
,
50 kHz
,
BW
,
1 kHz
on the functioning spectrum analyzer.
The signal on the functioning spectrum analyzer should be very stable
both in frequency and power as shown in
8.
Reconnect W32 A14J10.
Figure 8-7
100 MHz Narrow Span
If the 100 MHz signal is not measuring the correct power level, see the
“A16 Reference Assembly Troubleshooting”
section later in this chapter.