482 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area (Options 544, 550)
YTF Preselector
Removal
1.
. Remove cables W56 and W57 from the YTF
Preselector
(1)
.
2.
Remove the wire harness W11.
3.
. From the bottom of the instrument, remove the four
screws
(1)
). The YTF Preselector can now be removed from
the chassis (may require lifting up slightly on the low band switch).
Figure 16-48
YTF Preselector Removal
Table 16-13
Item
Keysight Part Number
W56
W57
Summary of Contents for N9030B