242 N9030B PXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A3 Digital IF Troubleshooting
12.
The analyzer should read 22.5 MHz at
−
25 dBm
±
Figure 7-8
22.5 MHz IF Input
13.
If the 22.5 MHz signal is not measuring the correct power level, see
Analog IF Troubleshooting” on page 230
in this service guide.
14.
If the 22.5 MHz signal is within tolerance, carefully reconnect the W42
cable to A3J19.