74 N9030B PXA Signal Analyzer Service Guide
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
Table 2-3
Initial Alignments
Alignment Description
Most Probable
Hard ware Failure
Related Hard ware
Main Loop YTO Pretune DAC
A14 Synthesizer
A16 Reference board is not
providing correct 100 MHz
or 4.8 GHz signals.
A20 YTO output power low.
Offset Loop VCO Pretune Alignment
A14 Synthesizer
A16 Reference board is not
providing correct 100 MHz
or 4.8 GHz signals
VCXO Tuning Curve
(Power up alignment only)
Measures the tuning voltage for the 100 MHz
VCO at a number of different frequencies. Uses
internal 10 MHz time base
A16 Reference
VCXO Loop Band wid th
Adjusts gain of the PLL loop amplifier. Uses the
10 MHz internal reference
A16 Reference
Second LO Tuning Curve
(Power up alignment only)
Measures the tune voltage for the 4800 MHz
VCO at a number of different frequencies. Uses
the 10 MHz internal reference
A16 Reference
Second LO Loop Band wid th
Adjusts gain of the PLL loop amplifier. Uses the
10 MHz internal reference.
A16 Reference
Ext Mix LO Alignment Algorithm
A13 Front End Assy
A15 Front End Controller
Dither Level Algorithm
(4 measurements)
A3 Digital IF
AIF LC Wide Prefilter Passband Tuning
Algorithm
Adjusts the LC prefilter centering with the
prefilter BW set to about 1.12 MHz.
Uses 322.5 MHz Comb Calibrator on the A-IF.
A2 Analog IF
AIF LC Wide Prefilter Passband Fine Tuning
Algorithm
A2 Analog IF
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...