544 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Front Frame Assembly
Front Panel Shield
Figure 16-102
Front Panel Shield
1.
. Remove ribbon cable W1.
2.
To remove the front panel shield
(1)
, remove the eight screws
(2)
(0515-0372).
3.
When replacing the front panel shield, torque all screws to 9 inch-pounds,
When replacing ribbon cable W1 press firmly on each side to ensure the
hooks catch on the boards mating connector.