N9030B PXA Signal Analyzer Service Guide 581
Functional Tests
Frequency Readout Accuracy
Procedure
1.
Configure the equipment as shown in
. Confirm the analyzer’s
built-in auto alignment has been performed within the past 24 hours.
2.
On the synthesized sweeper, press
PRESET
, then set the controls as
follows:
FREQUENCY
, 1505,
MHz
POWER LEVEL
, –10,
dBm
RF, On
3.
Set up the signal analyzer by pressing:
Mode
,
Spectrum Analyzer
Mode Preset
Input/Output, More
,
Freq Ref In
,
External
FREQ Channel
,
Center Freq
, 1505,
MHz
SPAN X Scale, Span, 2990, MHz
Trace/Detector
,
More
,
Detector
,
Sample
Single
4.
Press
Restart
.
Press
Peak Search
on the analyzer. If the instrument is functioning
correctly, the marker reading in the active function block will be between
. Record the marker value in the Marker
Frequency Readout column in
5.
On the signal analyzer, change the span and center frequency as listed in
6.
Change the synthesized sweeper frequency to match the center frequency
of the analyzer.
7.
Repeat
through
until the Marker Frequency Readout column
of
is complete.
Table 18-3
Frequency Readout Accuracy Results
Span
(MHz)
Center
Frequency
(MHz)
Minimum
Marker Frequency Readout
Maximum
2990
1505
1495.9 MHz
1514.1 MHz
127.2
1505
1504.56 MHz
1505.44 MHz
54.1
1505
1504.8122 MHz
1505.1878 MHz
7.95
1505
1504.97240 MHz
1505.0276 MHz
0.106
1505
1504.999630 MHz
1505.000370 MHz
1.98
517.59
517.58316 MHz
517.59684 MHz
1.98
832.50
832.49316 MHz
832.50684 MHz
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...