N9030B PXA Signal Analyzer Service Guide 269
LO Synthesizer/Reference Troubleshooting
A16 Reference Assembly Troubleshooting
Table 8-4
A16 Reference Assembly Signal Measurement Details
Signal
Description
Signal
Location
Expected
Frequency
(MHz)
Expected Power
Level
(dBm)
Special Instrument
Conditions
Figure
50 MHz
Ecal Out
A16J705
50 MHz
−
26 to
−
29 dBm
typical
While monitoring A16J705
output, place functioning
spectrum analyzer in Max
Hold. Press
System
,
Alignments
,
Align
Now
,
All
on PXA.
4.8 GHz
2nd LO Out
A16J702
4800 MHz
+11 dBm
Press
Freq
,
1 GHz
,
Span
,
Zero Span
on
PXA
4.8 MHz
1st LO Ref Out
A16J703
4800 MHz
+3 dBm
Press
Freq
,
50 MHz
,
Span
,
Zero Span
,
Meas Setup
,
PhNoise
Opt
,
Best Close in
PNoise
on PXA
50 MHz Ref Out A16J701
50 MHz
−
25 dBm
Press
Input/Output
,
RF
Calibrator
,
50 MHz
on
PXA
4.8 GHz Ref
Out
A16J701
4800 MHz
−
28 dBm
Press
Input/Output
,
RF
Calibrator
,
4.8 GHz
on
PXA
2.4 GHz Ref
Out
A16J718
2400 MHz
+10 dBm
300 MHz
LO AIF
A16J711
300 MHz
+10 dBm
100 MHz Ref
Output
A16J706
100 MHz
+13 dBm
100 MHz Ref B
Out
A16J716
100 MHz
+13 dBm
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...