210 N9030B PXA Signal Analyzer Service Guide
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Input Attenuator B Control Logic Verification (Option 503, 508, 513,
526)
Press
AMPTD
,
Attenuation 6 dB
on the analyzer and verify the 6 dB Step LED
DS17 is illuminated. Change to the input attenuation settings found in
and verify the proper LED's illuminate on the Front End Troubleshooting
board according to
If the LED's illuminate correctly, the switch control logic for the input
attenuators from the A15 Front End Control Assembly is correct. If the LED's
are not illuminating as expected, the most probable cause is the A15 Front End
Control Assembly.
Set the input attenuator back to 10 dB by pressing
AMPTD
,
Attenuation
,
10 dB
on the analyzer.
Table 6-7
Input Attenuator B LED’s
Attenuator
Setting (dB)
DS20
30 dB
Step LED
DS19
20 dB
Step LED
DS18
10 dB
Step LED
DS17
6 dB
Step LED
10
OFF
OFF
ON
OFF
16
OFF
OFF
ON
ON
20
OFF
ON
OFF
OFF
30
ON
OFF
OFF
OFF
40
ON
OFF
ON
OFF
50
ON
ON
OFF
OFF
60
ON
ON
ON
OFF
66
ON
ON
ON
ON
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...