N9030B PXA Signal Analyzer Service Guide 141
Instrument Messages
Condition Messages
Condition Errors 401 to 499, Power
This series of errors corresponds to the bits in the
STATus:QUEStionable:POWer register. The second column in the table shows
the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 409 indicates a 50 MHz Oscillator Unleveled condition has
been detected, error 1409 indicates that failure has been cleared.
This register is summarized as bit 3 of the STATus:QUEStionable register. See
section
“Condition Errors 601 to 699, Error Summaries” on page 143
.
Err#
Bit in
status
register
Message
Error or
Warning
More Information
401
0
RPP Tripped
W
(not currently in use)
403
1
Source Unleveled
W
405
2
Source LO Unleveled
E
(not currently in use)
407
3
LO Unleveled
E
(not currently in use)
409
4
unused
411
5
unused
413
6
unused
415
7
unused
417
8
unused
419
9
Preselector Overload
E
421
10
unused
423
11
unused
425
12
unused
427
13
unused
429
14
unused
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...