N9030B PXA Signal Analyzer Service Guide 571
Post-Repair Procedures
Post-Repair Procedures
Programming Model and Serial Numbers
Since the instrument model and serial numbers are stored only on the A7
Midplane board assembly, this information will be lost when this board
assembly is replaced. Until the instrument model and serial numbers are
restored the instrument application software will not allow any user
measurements to be made, due to the fact that all previously installed license
keys will not be accepted. However, the license files will not be automatically
deleted. Once the model and serial numbers are restore the license files will
then be recognized and accepted, returning the instrument to the capability
that was available prior to the A7 Midplane board assembly replacement.
To program the model and serial numbers into a replacement A7 Midplane
board assembly the N7814A Keysight X-Series Signal Analyzer Calibration
Application Software will be required.
Information regarding N7814A Keysight X-Series Signal Analyzer Calibration
Application Software can be found at:
http://www.keysight.com/find/calibrationsoftware
When the calibration software first detects the instrument it will recognize that
it does not have a valid model and serial number and will prompt the user to
enter the correct values.
Since the instrument model and serial numbers are stored on the A7 Midplane
board assembly, this should only be required when this assembly has been
replaced with a new one.
Be very careful when entering the instrument serial number into the
software, because once it has been written it can no longer be changed.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...