background image

H24.0.01.6C-07       page 1 of 4

Specification:

Measuring range:

0,00 to 14,00 pH

Resolution:

0,01 pH

Accuracy (device only):

± 0,02 pH  ±1 digit  

(at nominal temperature = 25°C)

pH-electrode:

pH-electrode GE014 (standard electrode, incl. in scope of supply),
plug in via front-side electrode socket (Cinch).
GE 014: combined measuring and reference electrode with refillable 3 mol-KCl-electrolyte.

measuring range: 2 to 12 pH, temperature: 5 to 45°C, conductivity: >200 µS/cm.

For permanent measurements in a highly alkaline environment, special electrodes have to
be used. (Please refer to special accessories)

Input resistance:

approx. 10

12 

Ohm

Display:

approx. 13 mm high, 3½ digit LCD

Working temperature:

0 to 45°C

Relative humidity:

0 to 80 % r.F. (non condensing)

Power supply:

9V-battery Type IEC 6F22 (included in scope of supply)

Battery service life:

approx. 200 h

Low battery warning:

Automatic; "BAT" displayed in case of low battery voltage

Dimensions:

approx. 106 x 67 x 30 mm (H x W x D)

Weight:

approx. 200g (incl. battery and electrode GE014)

Calibrating:

3 turning knobs for:
1. temperature compensation 0 to 90°C
2. pH 7-value
3. pH X-value (e.g. pH 1,09, pH 4, pH 10 or pH 12, depending on working environment)

EMC:

The device corresponds to the essential protection ratings established in the Regulations
of the Council for the Approximation of Legislation for the member countries regarding
electromagnetic compatibility (89/336/EWG).
additional errorr: <1%

Operating Manual for Digital-pH-Meter

GPH 014

Disposal instruction:

The device must not be disposed in the unsorted municipal waste! Send the device directly to us (sufficiently stamped), if it
should be disposed. We will dispose the device appropriate and environmentally sound.

Points to be observed during operation

a.) In case of low battery voltage "BAT" is displayed; make sure to exchange battery immediately as too low an operation

voltage may lead to incorrect measuring results.
We recommend to take out battery if device will not be operated for some time or storaged over 50°C.

Please note: If the battery voltage falls even lower the voltage may not be sufficient for "BAT" to be displayed so that

there will be no "BAT" indication although the battery has been used up.
We recommend to make it a rule to always check the battery if the values indicated seem to be
completely out of range.

b.) Make sure to maintain device an electrode properly and to operate it in accordance with the specification listed  (do not throw,

knock etc.).

c.) The electrode must be stored in 3 mol/l KCl (ours type: KCl3M), in order to avoid drying out (excluded GE103).
d.) Before measuring the protection-cap has to be removed.
e.) Our pH-electrodes can be used in a degree range from 90° to ±45° compared to horizontal

D - 93128 Regenstauf, Hans-Sachs-Straße 26

Fon: 0049-9402 / 9383-0
Fax: 0049-9402 / 9383-33

GREISINGER 

electronic GmbH

Summary of Contents for GPH 014

Page 1: ...egislation for the member countries regarding electromagnetic compatibility 89 336 EWG additional errorr 1 Operating Manual for Digital pH Meter GPH 014 Disposal instruction The device must not be disposed in the unsorted municipal waste Send the device directly to us sufficiently stamped if it should be disposed We will dispose the device appropriate and environmentally sound Points to be observe...

Page 2: ...x 20 to 30 seconds for display to show stable value then use turning knob at the far right side pH 7 to set value to 7 00 Setting of the second calibration point Rinse electrode with distilled water once again dry electrode place it in buffer solution pH 4 0 Wait for another 20 to 30 seconds for display to show stable value then use turning knob at the far left side pH X to set value to 4 00 Safet...

Page 3: ...Use prick lance first Points to be observed during measuring As resistance values resulting from both the pH electrode and from the measuring device are generally very high when conducting pH measurements the electrode should not be hand held but put down during the measuring operation in order to avoid a shift in voltage parameters between the electrode and the measuring device In case there is n...

Page 4: ... if necessary top up with 3 mol l KCl solution through filling hole using a syringe or pipette 2 2 Crystaization of the 3 mol l KCl is not avoidable Crystalized KCl at protection cap or closing collar can easily be removed by fingernail or cloth and is no defect or reason for reclamation 2 3 Dirty electrodes have to be cleaned You will find suitable cleaning agents for the pH glass diaphragm in th...

Reviews: