N9030B PXA Signal Analyzer Service Guide 461
Assembly Replacement Procedures
RF Area (Options 503, 508, 513, 526)
2.
Refer to
. Place the second switch/bracket assembly
(1)/(2)
on the bracket previously installed. Assure label on switch is facing up. To
attach use the two screws
(3)
(0515-1992).
Figure 16-29
Switch 2 Placement
3.
Secure the bracket to the Low Band Switch bracket with the single screw
(4)
). Do not torque.
4.
Align switches so they are parallel with side of instrument then torque the
four screws
(3)
securing the switches to the brackets to 6 in lbs. Torque
the screw
(4)
that secures the switch brackets together to 9 inch-lbs.