N9030B PXA Signal Analyzer Service Guide 67
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
Instrument Hangs at the Keysight Splash Screen
A problem of the instrument hanging at the Keysight splash screen could be
caused by many different things. It could be due to a down power supply, a
processor hardware problem, an instrument boot-up process error, etc.
1.
Remove the AC power cord and then remove the instrument cover. Refer
Chapter 16, “Assembly Replacement Procedures”, on page 429
in this
manual.
2.
Remove the Top Brace. Refer to
Chapter 16, “Assembly Replacement
in this manual.
3.
With the AC power applied and the On/Off button turned on, verify that all
of the power supply voltages are at their proper level. This can easily be
done by viewing the power supply LEDs on the back side of the A7
Midplane board. See the
"A7 Midplane Board Assembly Troubleshooting"
.
Are all of the power supply voltages at the proper level?
If yes:
After verifying that the connections from the A7 Midplane board to
the A4 Processor board are not at fault, replace the A4 Processor
board.
If not:
After verifying that the connections between the A6 Power Supply,
the A7 Midplane board, and the A8 Motherboard are all
mechanically and electrically secure, replace the A6 Power Supply
assembly.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...