A25 Wideband Analog IF Troubleshooting - Option B1X and B85
A25 Wideband Troubleshooting - Option B1X and B85
Verify the 300 MHz IF Input level
1.
Turn the instrument off.
2.
Remove the instrument cover. Refer to
3.
Turn on the signal analyzer and wait for the instrument to complete the
boot up process.
4.
Press
System
,
Alignments
,
Auto Align
,
Off
.
5.
Press
Mode
,
IQ Analyzer
6.
Press
Meas
,
Complex Spectrum
7.
Press
Input/Output
,
RF Calibrator
,
50 MHz
8.
Press
Mode Setup
,
IF Path
,
160 MHz
(or
85 MHz
)
9.
Verify the 50 MHz signal is at
−
25 dBm on the analyzer display by pressing
FREQ
,
50 MHz
,
SPAN
,
1 MHz
,
Peak Search
on the analyzer. The marker
readout should be 50 MHz at
−
25 dBm
±
3 dB. If this reference signal is