N9030B PXA Signal Analyzer Service Guide 231
Analog/Digital IF Troubleshooting
A2 Analog IF Troubleshooting
8.
Carefully disconnect W25 cable at A15J900 as shown in
Figure 7-1
A15 RF Front End Control Assembly
9.
Connect A15J900 to a functioning spectrum analyzer using the
appropriate MMCX connectors and cables.
10.
Press
Freq
,
322.5 MHz
,
Span,
1 MHz
,
Peak Search
on the functioning
spectrum analyzer.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...