96 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–300 to –399, Device-Specific Errors
–440
Query UNTERMINATED after
indefinite response
A query was received in the same program message after a query
requesting an indefinite response was executed.
Err#
Message
Verbose/Correction Information
–300
Device-specific error
An instrument error occurred and the exact problem cannot be specifically
identified. Report this error to the nearest Keysight Technologies sales or
service office.
–310
System error;
An internal system-type error has occurred. The exact problem cannot be
specifically identified. Report this error to the nearest Keysight
Technologies sales or service office.
–310
System error; A
license will soon
expire;<feature code>
will expire in <time>
The indicated feature/software will expire in the specified time. Contact
Keysight Technologies to purchase continued use of this functionality.
–310
System Error; enable
GPIB controller mode
Press System, I/O Config, GPIB and set GPIB Controller to Enabled so that
the analyzer can control the source over GPIB
–310
System error; Error
transmitting a LAN
event to the network.
Communication with the network driver failed.
–310
System error; Failed
to initialize the PTP
clock to current
time.
Failure communicating with the DMC libraries’ PTP controller.
–310
System error; Failed
to instantiate the
PTP ordinary clock.
Failure in the starting up the DMC libraries’ PTP controller.
–310
System error; feature
<feature code> not
licensed
The specified feature, for example “N9073A-TR2” is not licensed. The
license may have expired. You cannot use it until you get a license.
–310
System error; Feature
expired; <feature
code>
The specified feature has expired. The license is no longer valid.
–310
System error; License
installation
failed;<feature code>
The license installation of the specified feature, for example
“N9073A-TR2”, has failed. You should refer to the event log in the control
panel for more details.
–310
System error; License
removal failed;
<feature code>
The license removal of the specified feature, for example “N9073A-TR2”
has failed. You should refer to the event log in the control panel for more
details.
Err#
Message
Verbose/Correction Information
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...