N9030B PXA Signal Analyzer Service Guide 63
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
Fan(s) Are Not Operating
Control of the instrument fans comes from the A6 Power Supply assembly. This
signal is routed from the A6 Power Supply through the A7 Midplane board,
where there is a test point and LED to monitor the level, and is then routed to
the A8 Motherboard where it is filtered before being sent to the Fans. When the
Fans do not come on it could be due to any one of these assemblies. This
procedure will help to determine which one is the cause.
This procedure assumes that the green Power On LED on the front panel does
turn on when the instrument is turned on. If it doesn't, refer to the
Power On LED Does Not Illuminate”
section before proceeding.
1.
The instrument fans will only turn on when the instrument is connected to
an AC power source that has a voltage level and frequency of that
specified for the instrument and the front panel On/Off button has been
pressed. Before proceeding verify that these requirements are being met.
Refer to the instrument rear panel for these requirements.
2.
Remove the AC power cord and then remove the instrument cover. Refer
Chapter 16, “Assembly Replacement Procedures”, on page 429
in this
manual.
3.
Are all three fans not spinning?
If yes:
Proceed to
.
If not:
Proceed to
.
4.
Remove the Top Brace. Refer to
Chapter 16, “Assembly Replacement
in this manual.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...