590 N9030B PXA Signal Analyzer Service Guide
Functional Tests
Frequency Response (Flatness)
Figure 18-5
Frequency Response Test Setup
Procedure
1.
Zero and calibrate the power meter and power sensor as described in the
power meter operation manual.
2.
Configure the equipment as shown in
.
3.
Assure the signal analyzer’s built-in auto alignment has been performed
within the last 24 hours.
Power Splitter
(Opt 503, 508, 513, 526)
Frequency Range
: 50 MHz to 26.5 GHz
Tracking between ports
: < 0.25 dB
11667B
Power Splitter
(Opt 544, 550)
Frequency Range
: 50 MHz to 50 GHz
Tracking between ports
: < 0.40 dB
11667C
Synthesized Sweeper
Frequency Rang
e: 50 MHz to 50 GHz
a
PSG
a. Maximum frequency of the sweeper must be greater than or equal to maximum specified fre-
quency of the PXA.
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Connect the power splitter to the signal analyzer input using the
appropriate adapter. Do not use a cable. An adapter will only be necessary
if the PXA input connector is Type-N.