N9030B PXA Signal Analyzer Service Guide 585
Functional Tests
Amplitude Accuracy at 50 MHz
Amplitude Accuracy at 50 MHz
Test Limits
Amplitude Accuracy should remain within 1.13 dB of the measured source
value across the range of source levels and changes in resolution bandwidth.
The preamp (Option P03, P08, P13, P26) should remain within ±1.3 dB of
measured values.
A synthesized sweeper is used as the signal source for the test. The source
amplitude is varied using the signal source amplitude control. The attenuation
and resolution bandwidth are varied on the signal analyzer. The source
amplitude is measured by the power meter and signal analyzer at each setting,
and the values compared. The difference between each pair of measurements
indicates the amplitude accuracy.
Item
Critical Specifications
Recommended
Keysight Model
Adapter
Type-N (m), to 3.5 mm (m)
1250-1743
Adapter
3.5 mm (f) to 3.5 mm (f)
83059B
Attenuator, 30 dB
Accuracy
: < 0.5 dB at 50 MHz
11708A
Cable
3.5 mm, 1 meter
11500E
Cable
BNC, 120 cm
10503A
Power Meter
Compatible with power sensor
E4418B
Power Sensor
Amplitude Range
: -25 dBm to 10 dBm
8485A
Power Splitter
3.5 mm (f) connectors
6 dB loss
11667B
Synthesized Sweeper
Typical Temperature Stability
: 0.01 dBc/°C
PSG
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...