30 N9030B PXA Signal Analyzer Service Guide
Overview
Service Equipment You Will Need
Front End Controller Troubleshooting Kit
The Front End Troubleshooting kit contains a PC board, required cables, and
installation instructions to verify the switch control logic from the A15 Front
End Control board to the lower level front end components is correct.
Oftentimes when troubleshooting an RF front end problem, the logic needs to
be verified before one of the front end components is changed. This
troubleshooting kit will help identify the defective area in a timely manner. In
order to effectively isolate A15 Front End Control board problems from front
end component problems, it is highly recommended that the N9020-60005
Front End Troubleshooting kit is purchased.
The Front End Troubleshooting kit lower level items can be purchased
individually or as a complete kit with instructions. The complete listing of kit
components, descriptions, and part numbers can be found below.
Keysight Part
Keysight Part Number
Notes
Front End Troubleshooting kit
N9020-60005
Front End Troubleshooting board
E4410-60115
Part of N9020-60005 Troubleshooting kit
Cable, Low Band switch
E4410-60160
Part of N9020-60005 Troubleshooting kit
Cable, RF Downconverter
E4410-60156
Part of N9020-60005 Troubleshooting kit
Cable, YTF Preselector
E4410-60158
Part of N9020-60005 Troubleshooting kit
Cable, Input Attenuators
E4410-60157
Part of N9020-60005 Troubleshooting kit
Cable, Troubleshooting
8121-1400
Right-angle sma (m) to
right-angle mmcx (m)
(must be ordered separately)
Connector, MMCX (f) to SMA (f)
n/a
www.hubersuhner.com
item number: 31_MMCX-SMZ-50-1/111_OE
(must be ordered separately)
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...