N9030B PXA Signal Analyzer Service Guide 535
Assembly Replacement Procedures
Fan Assembly
Replacing a Fan
1.
. To replace a fan, it is necessary to remove the 4
plastic rivets
(3)
) that attach the fan
(1)
and guard
(2)
to the
fan bracket. To do this use a small screwdriver to unscrew the rivet and
remove it from the bracket. The rivets may need to be cut or pushed out
from behind. They are not reusable.
Figure 16-96
Fan Replacement
2.
Position the new fan on the work surface with the arrows pointing to the
right and down.
3.
Orient the fan guard so that the rings are on the outside, and that the “V”
of the support legs are pointing to the neighboring fan.
4.
With the rivets center posts raised, snap the rivets into place through the
fan grill and fan bracket and into the fan. Screw the center posts into
place. This may require a small hammer or push tool to push the rivets in.