N9030B PXA Signal Analyzer Service Guide 233
Analog/Digital IF Troubleshooting
A2 Analog IF Troubleshooting
Verifying the 22.5 MHz Output Power
1.
Perform an instrument shutdown.
2.
Turn the instrument over so the bottom side of the analyzer is facing up.
3.
Turn on the PXA Signal Analyzer and wait for the instrument to complete
the boot up process.
4.
Press
System
,
Alignments
,
Auto Align
,
Off
.
5.
Press
Input/Output
,
RF calibrator
,
50 MHz
.
6.
Verify the 50 MHz signal is at
−
25 dBm by pressing
FREQ
,
50 MHz
,
SPAN
,
1 MHz
,
Peak Search
on the analyzer. The marker readout should be 50
MHz at
−
25 dBm
±
3 dB. If this reference signal is measuring incorrectly,
see
Chapter 4, “RF Section Troubleshooting (RF/Microwave Analyzers)”
this service guide.
7.
To continue verifying press
SPAN
,
Zero Span
. Verify the input attenuator
on the PXA is set to 10 dB. Look near the top of the display near the center
and verify that
Atten: 10 dB
is visible. If needed change the input
attenuator by pressing
AMPTD
,
Attenuation
,
10 dB
on the analyzer.
8.
, carefully disconnect the W42 cable at A3J19.
Figure 7-3
A3 Digital IF Cables
9.
Connect the W42 cable to the MMCX female to SMA female connector.
Use an appropriate cable to go from the SMA connector to the RF input of
a functioning spectrum analyzer to verify the 22.5 MHz I.F. and amplitude
is correct.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...