532 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Motherboard Assembly
5.
Replace the Midplane assembly. Refer to the
replacement procedure.
6.
Replace the Front End Control assembly. Refer to the
replacement procedure.
7.
Replace the LO Synthesizer assembly. Refer to the
replacement procedure.
8.
Replace the AIF/DIF assembly. Refer to the
replacement
procedure.
9.
Replace the CPU assembly. Refer to the
replacement
procedure.
10.
Replace the power supply assembly. Refer to the
replacement procedure.
11.
Replace the reference assembly. Refer to the
replacement procedure.
12.
Replace the rear panel. Refer to the
replacement procedure.
13.
If Option B1X is present in the instrument, install the two option boards.
14.
Ensure all cables are plugged in or connected and torqued.
15.
Replace the RF bracket. Refer to
. Torque the
screws to 9 inch-pounds.
16.
Replace the Front Frame Assembly. Refer to the
replacement procedure.
17.
Replace the top brace and reference bracket. Refer to the
replacement procedure.
18.
Replace the instrument outer case. Refer to the
replacement procedure.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...