N9030B PXA Signal Analyzer Service Guide 477
Assembly Replacement Procedures
RF Area (Options 544, 550)
Replacement
1.
. Position the attenuator and magnetic shield in the
bracket so that the attenuator ribbon connector end is “up”.
2.
Replace the two screws (
) that attach the attenuator to the
bracket. Torque to 9 inch-pounds.
3.
Position the attenuator and bracket into the chassis and attach with the
two screws (
). Torque to 9 inch-pounds.
4.
Replace the ribbon cable and semi-rigid cables W2, W3, W4, or W6 to the
attenuator. Torque the semi-rigid cables to 10 inch-pounds.
5.
. Position the RF bracket onto the chassis and
replace the sixteen screws (
). Torque to 9 inch-pounds.
6.
Replace the front panel. Refer to the
replacement
procedure.
7.
Replace the instrument outer case. Refer to the
replacement procedure.