N9030B PXA Signal Analyzer Service Guide 557
Post-Repair Procedures
Post-Repair Procedures
Additional Tasks
Calibration Data Backup and Restore
In order for the instrument being repaired to not need all of the instrument
adjustments and performance verification tests to be run after the A5 Disk
Drive is replaced the calibration data will need to be backed up onto an
external drive prior to removing the existing disk drive.
Data Backup
There are two different backup procedures that might need to be followed.
Which backup procedure to be run will depend on whether the software will
run or not.
If the Instrument Software Application Will Run
1.
Connect a USB keyboard and mouse to two of the instrument rear
panel USB ports
2.
Connect a USB FLASH drive to one of the front panel USB port.
3.
Press
System
,
Alignments
,
Backup or Restore Align Data
.
4.
The Alignment Data Wizard window as shown in
will
appear.
While the backing up and restoring of the calibration data to the new disk
drive will return the instrument performance to the state that it was prior
to a repair this does not guarantee that all instrument performance
parameters meet their specified values. A full instrument calibration would
be required to verify this.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...