N9030B PXA Signal Analyzer Service Guide 337
Hardware Options
A26 Wideband Digital IF Troubleshooting - Option B2X and B5X
A26 Wideband Digital IF Troubleshooting - Option B2X and B5X
Verify the Wideband Signal Path
Normally failure of this assembly will occur during extensive alignments. Verify
if there are alignment error related to the WBDIF located in the
E:\AlignDataStore\AlignmentHistrory.txt files.
If errors are found related to WBDIF replace the A26 Wideband Digital IF
assembly. To troubleshoot this module set the instrument up as below:
1.
Turn on the analyzer and wait for the instrument to complete the boot up
process.
2.
Press
System
,
Alignments
,
Auto Align
,
Off
.
3.
Press
Mode
,
IQ Analyzer
4.
Press
Meas
,
Complex Spectrum
5.
Press
Input/Output
,
RF Calibrator
,
50 MHz
6.
Press
Span
,
10 MHz
7.
Press
Mode Setup
,
IF Path
,
255 MHz
or
510 MHz
8.
Verify the 50 MHz signal is at
−
25 dBm on the analyzer display by pressing
Peak Search
on the analyzer. The marker readout should be 50 MHz at
−
25 dBm
±
3 dB.
9.
If this signal is not present, determine if the issue is only present with the
wideband path. Press
Mode Setup
,
IF Path
,
40 MHz
. If the signal is now
present, there is a likely failure of the A26/A27 Wideband Digital IF or the
A25 Wideband Analog IF
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...