218 N9030B PXA Signal Analyzer Service Guide
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Oscilloscope Test
Measurements can be made to verify the current logic is getting to the
appropriate switch from the A15. In order to perform this measurement, the
outer cover and chassis RF bracket on the right hand side of the instrument
must be removed. Refer to
Chapter 16, “Assembly Replacement Procedures”,
for the removal procedures.
illustrate the measurement location
where the oscilloscope probe should be used to measure the logic. There are
two connector bodies with three wires each. On one connector, the three wires
are green, black and orange. On the other connector, the three wires are green,
black, and yellow. On each connector, black is ground.
On analyzers with Option 508, 513, and 526, both connectors are used; one
connects to SW1 and the other to SW2. It does not matter which connector
connects to which switch. Both connectors are programmed to behave
identically.
However, on analyzers with Option 544 and 550, only the green, black, yellow
connector is used to control SW4.
Figure 6-5
SW1 and SW2 Connector Location (Option 508, 513, and 526)
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...