262 N9030B PXA Signal Analyzer Service Guide
LO Synthesizer/Reference Troubleshooting
A14 LO Synthesizer and A20 YTO Troubleshooting
Verifying the 1st LO Power and Frequency:
1.
Turn on the N9030B Signal Analyzer and wait for the instrument to
complete the boot up process.
2.
Press
FREQ
,
50 MHz
,
SPAN
,
Zero Span
.
3.
. Remove W21 from both A13A1J5 and A14J20.
4.
Disconnect W20 from A13A1J4.
5.
Refer to
. Connect the input of an 87300B directional coupler to
the end of W20 that was disconnected from A13A1J4, using a SMA (m-f)
elbow. Connect a short SMA cable from one output of the directional
coupler to A14J20. Connect another cable from the coupled output of the
directional coupler to the input of the functioning spectrum analyzer.
These connections bypass the LO distribution amplifiers on the A13A1 RF
Front End Assembly.
Figure 8-8
Directional Coupler Connection
6.
Set the functioning spectrum analyzer to a center frequency of 5172.5
MHz and a span of 5 MHz. Set the reference level to +10 dBm.
7.
Press
Peak Search
on the functioning spectrum analyzer. If the functioning
spectrum analyzer has a preselector peak or preselector centering
function, perform that operation to guarantee that the preselector is
optimized for the measurement.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...