N9030B PXA Signal Analyzer Service Guide 77
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
AIF 140 Variable Attenuator Algorithm
Ref power at atten 12 dB
(Option B1X)
A25 Wide band A-IF
DIF 40 Variable Attenuator Algorithm
Ref power at atten 21 dB
A3 Digital IF
Narrow Band Step Cal Ad justment Alignment
Algorithm
Alignment signal
originates on A3
D-IF, then goes
through A16
Reference through
the input
attenuators,
through A13 Front
End, through A2
Analog IF, and back
to A3 Digital IF. If
Option B1X present,
A25 WB A-IF
switches calibrator
signal to A16
Reference.
Narrow Band Step Cal Alignment Algorithm
A16 Reference
A9/A10 Attenuators
DIF 40 Step Cal Adjustment Alignment
Algorithm
Since this is wide
band alignment
done for 40 MHz
BW signal path, A2
Analog IF is not
included in the
signal path.
DIF 40 Step Cal Alignment Algorithm
AIF 140 Step Cal Adjustment Alignment
Algorithm
(Option B1X)
A25 Wide band A-IF
AIF 140 Step Cal Alignment Algorithm
Table 2-3
Initial Alignments
Alignment Description
Most Probable
Hard ware Failure
Related Hard ware
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...