92 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Advisory Messages
Reading SNS data…
The Keysight Smart Noise Source has been connected and the application
is reading the device EEPROM data. Please wait until complete before
continuing.
Recalled File <filename>
A file recall (open/load) was successfully completed.
Refer to online help for
assistance with DSB
measurements
The Double Side Band measurement requires careful setup to obtain valid
results. Please refer to the manuals for help with this setup.
Requested timeslot number is
not present.
The selected timeslot is not on. (Timeslot is referenced to the trigger
point.)
RF Start freq should be
greater than IF Fixed freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF Start freq should be greater than the IF Fixed
freq for a DSB (for DSB measurements the setup uses LSB values)
downconverter swept LO setup. Use the graph icon on the DUT setup form
to clarify the setup required.
RF Start freq should be
greater than IF start freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF start freq should be greater than the IF Start
freq’s for an LSB downconverter fixed LO setup. Use the graph icon on the
DUT setup form to clarify the setup required.
RF Start freq should be
greater than LO fixed freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF start freq should be greater than the LO fixed
freq’s for an USB downconverter fixed LO setup. Use the graph icon on the
DUT setup form to clarify the setup required.
RF Start freq should be
greater than LO Start freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF start freq should be greater than the LO Start
freq’s for an USB downconverter swept LO setup. Use the graph icon on
the DUT setup form to clarify the setup required.
RF Stop freq should be
greater than IF Stop freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF Stop freq should be greater than the IF Stop
freq’s for an USB or DSB (for DSB measurements the setup uses LSB
values) downconverter fixed LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
Scale/Div only applies in Log
Y Scale
Setting the Scale/Division only makes sense when you are using a
logarithmic Y scale.
Signal Track is turned off
when Zero Span is selected
Signal Track is not available when you have selected Zero Span. So if Zero
Span is entered while in Signal Track is On, Signal Track is turned off.
Signals deleted
The signals in the signal list were successfully deleted.
Message
Description/Correction Information
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...