52 N9030B PXA Signal Analyzer Service Guide
Boot Up and Initialization Troubleshooting
Check the Basics
Check the Basics
Before calling Keysight Technologies or returning the instrument for service,
please make the following checks:
1.
Is there power at the power outlet? At the power receptacle on the
instrument?
2.
Is the instrument turned on? Check to see if the front panel LED is green,
which indicates the power supply is on.
3.
If other equipment, cables, and connectors are being used with the
instrument, make sure they are connected properly.
4.
Review the procedure for the measurement being performed when the
problem appeared. Are all the settings correct?
5.
If the instrument is not functioning as expected, return the unit to a known
state by pressing the
Mode Preset
key.
6.
Is the measurement being performed, and the results that are expected,
within the specifications and capabilities of the instrument? Refer to the
Specifications Guide for specifications.
7.
In order to meet specifications, the instrument must be aligned. Press
System
,
Alignments
,
Align Now
,
All
. The diagnostic tests should all pass.
If the instrument displays a failure during these tests, refer to
.
8.
Check to see if the instrument has the latest firmware before starting the
troubleshooting procedure. Press
System
,
Show System
. The firmware
revision is listed under
Firmware Revision
. For more information, refer to
Chapter 19, “Instrument Software”, on page 609
9.
Is the instrument displaying an error message? If so, refer to
“Instrument Messages”, on page 83
for more information.
10.
If the necessary test equipment is available, perform the functional checks
in
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...