588 N9030B PXA Signal Analyzer Service Guide
Functional Tests
Amplitude Accuracy at 50 MHz
Testing Preamp Option (P03, P08, P13, P26)
Instruments containing Option P03, P08, P13, P26 must have the preamp
function turned on and tested.
Procedure
1.
On the analyzer, press
AMPTD Y Scale
,
More
,
Internal Preamp
,
Low Band
.
2.
Connect the equipment as shown in
, using a
30 dB pad between the adaptor and the signal analyzer input.
3.
Set the synthesized sweeper amplitude to the value listed in the Nominal
.
4.
Set the signal analyzer input attenuation to 0 dB.
5.
Set the Span as listed in
.
6.
Record the synthesized sweeper amplitude, as measured by the power
meter, in the Power Meter Amplitude column of
7.
On the signal analyzer, press
Restart
.
8.
Wait for the analyzer to finish averaging.
9.
Press
Peak Search
.
10.
Record the signal amplitude as measured by the analyzer in the measured
amplitude column of
11.
Calculate the signal amplitude accuracy using the following equation:
Amplitude Accuracy Error = Me 30 dB - Power_meter
12.
Record the results under the Amplitude Accuracy Error column of
Table 18-6
Amplitude Accuracy Results (Option P03, P08, P13, P26)
Nominal
Source
Amplitude
(dBm)
Low- band
Preamp
Res BW
(kHz)
Span
(kHz)
Measured
Amplitude
Meas_amp
(dBm)
Power Meter
Amplitude
Power_meter
(dBm)
Amplitude
Accuracy
Error
(dB)
Test Limit
(dB)
-13
ON
1
106
±1.30 dB
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...