N9030B PXA Signal Analyzer Service Guide 583
Functional Tests
Second Harmonic Distortion (SHD)
Figure 18-3
Second Harmonic Distortion Test Setup
Procedure
1.
Configure the equipment as shown in
.
2.
Press
Mode
,
Spectrum Analyzer
,
Mode Preset
on the signal analyzer and
Preset the synthesized sweeper.
3.
Set up the synthesized sweeper by pressing:
Frequency
, 40,
MHz
Amplitude
, –5,
dBm
4.
Set up the signal analyzer by pressing:
Input/Output
,
More
,
Freq Ref In
,
External
FREQ Channel
,
Center Freq
, 40,
MHz
SPAN X Scale
,
Span
, 1,
MHz
AMPTD Y Scale, Attenuation, Mech Atten
, 10,
dB
5.
On the analyzer, press
Peak Search
.
6.
Adjust the synthesized sweeper amplitude for a signal analyzer display of
–5 dBm ±0.1 dB.
7.
On the signal analyzer, activate the marker delta function by pressing the
Marker
and
Delta
keys.
8.
On the signal analyzer, press:
FREQ Channel
,
Center Freq
, 80,
MHz
Meas Setup
,
Average/Hold Number
, 20,
Enter
Trace/Detector, Trace Average
Single